![Primes BeamMonitor BM+ 100S Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/primes/beammonitor-bm-100s/beammonitor-bm-100s_original-instructions-manual_1601729019.webp)
BeamMonitor BM+
35,0(6
19
Revision 02 EN - 02/2022
8.2
Power supply (Power In)
The BeamMonitor BM+ requires a voltage supply of 24 V ± 5 % (DC) for the operation. A suitable power sup-
ply is included in the scope of delivery.
BeamMonitor BM+ 100S
PRIMES power supply
Fig. 8.3:
Connection of the power supply using the BM+ 100S as an example
Harting M12-P-PCB-THR-2PC-5P-LCOD-M-STR
FE
4
1
2
3
Pin
Function
1
+24 V
2
Not assigned
3
GND
4
Not assigned
5
FE (functional earth)
Tab. 8.1:
Pin assignment of the connection socket for the PRIMES power supply