BeamMonitor BM+
35,0(6
21
Revision 02 EN - 02/2022
8.5
Connection of the BeamMonitor BM+ and PowerMonitor PM 48/100 to the PC
For sufficient absorption of the radiation behind the measurement zone, you can use the PRIMES laser
power meter PM 48/100. The water-cooled PM 48/100 will measure the current laser power and provide ad-
ditional information on the flow rate and temperature of the cooling water.
NOTICE
Damage / Destruction of the device due to overvoltage
When disconnecting the electrical lines during operation (with the supply voltage applied), voltage
peaks occur which can destroy the communication modules of the devices.
X
Disconnect the supply voltage from the device before disconnecting the bus cables.
When connecting several devices, always use only one PRIMES power supply unit for the voltage
supply.
BeamMonitor BM+ 100S
PowerMonitor PM 100
Ethernet
PC
PRIMES power supply
L
Max
= 10 m
PRIMES bus
RS485
Crossover cable
Fig. 8.5:
Connection to the PC using the example of the BM+ 100S together with the PowerMonitor PM 100