16.3.3
Assemble the detector
Only use isolating plastic screws to fasten the detector to prevent any electric noise signals from
being interspersed. Do not forget the foam rubber spacer during installation, otherwise the rota-
tional disc could be mechanically blocked by the screws. The foam rubber spacer also ensures
mechanical decoupling of the detector.
1. Place the foam rubber spacer (C) on the mounting surface of the detector (see Fig. 16.5 on page 67).
2. Connect the cables.
NOTICE
Blocking the rotational disc
If the screws are tightened too firmly, they might block the rotational disc!
X
Only tighten the screws hand-tight. The foam rubber spacer must only be compressed to no more
than half its original thickness.
3. Fasten the detector with the two plastic screws (D).
C
D
Fig. 16.5:
Assembling a new detector
16.3.4
Replace cover
1. Place the cover on the casing (see Fig. 16.1 on page 65).
2. Screw the cover on tight with the four Torx screws T8.
3. Check that the cover is securely seated.
The cover must lay flush with the casing.
Содержание BeamMonitor BM+ 100S
Страница 2: ......
Страница 3: ...BeamMonitor BM 35 0 6 3 Revision 02 EN 02 2022 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Страница 58: ...BeamMonitor BM 35 0 6 58 Revision 02 EN 02 2022 13 Declaration of conformity...
Страница 59: ...BeamMonitor BM 35 0 6 59 Revision 02 EN 02 2022...
Страница 68: ...BeamMonitor BM 35 0 6 68 Revision 02 EN 02 2022...