Chapter C - SERVICE MODE
Planmeca Compact i C-47
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n.301 Headrest software major version
Shows headrest software major version which is in use on the headrest PCB. When in Set
mode, shows headrest software major version on the MCB software chip that can be
upgraded to headrest PBC.
n.302 Headrest software minor version
Shows headrest software minor version which is in use on the headrest PCB. When in Set
mode, shows headrest software minor version on the MCB software chip that can be
upgraded to headrest PBC.
n.303 Upgrade headrest software
Press S key to upgrade headrest software on the headrest PCB into version that is on the
MCB software chip.
n.304 Headrest manual angle rotation speed
In this service mode the headrest manual angle rotation speed is adjusted. Minimum value =
0, Maximum = 15. Default value = 6. Larger value is faster.
n.305 Headrest manual height movement speed
In this service mode the headrest manual height movement speed is adjusted. Minimum
value = 0, Maximum = 15. Default value = 6. Larger value is faster.
n.306 Headrest manual length movement speed
In this service mode the headrest manual length movement speed is adjusted. Minimum
value = 0, Maximum = 15. Default value = 10. Larger value is faster.
n.307 Headrest automatic movement speed
In this service mode the headrest automatic movement speed is adjusted. Minimum value =
0, Maximum = 15. Default value = 6. Larger value is faster.
n.308 Headrest length current limit value
In this service mode the headrest length current limit value is adjusted. Minimum value = 0,
Maximum = 15. Default value = 11. Larger value indicates that more current (force) is needed
to stop.
n.309 Headrest joint 2 closing gap stopping term value
In this service mode the headrest joint 2 closing gap stopping term value is adjusted. Mini-
mum value = 0, Maximum = 15. Default = 0. Larger value indicates that headrest stops less
easy, might be needed to increase when using faster speeds.
n.310 Headrest joint 2 (near cushion) ADC-value
Value is typically from 528 to 908.
After pressing the S key, the joint can be driven.
This service mode is used when the headrest upholstery is replaced.
n.311 Headrest joint 1 (near backrest) ADC-value
Value is typically from 85 to 477.
After pressing the S key, the joint can be driven.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......