Chapter G - PARTS REPLACEMENT & REPAIR
Planmeca Compact i G-39
VALVES & FILTERS REPLACEMENT
Technical Manual
f)
Adjust the pressure regulator as described in section 3.1 “Adjusting internal water & air
pressures” on page F-26:
•
water 280 kPa (2.8 bar) (41 psi)
•
air 550 kPa (5.5 bar) (80 psi)
g)
Apply pressure to test that there are no air or water leaks.
7.5 Replacing the water / air regulator membrane
a)
Remove the lock ring from the regulator cap.
b)
Unscrew the four attachment screws of the regulator cap and remove the cap.
c)
Replace the membrane. Note, that the air regulator membrane has a small opening in the
middle of the membrane.
Lock ring
Regulator cap
Membrane
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......