Chapter D - HELP & ERROR MESSAGES
Planmeca Compact i D-5
HELP MESSAGES
Technical Manual
HE 50
No glass in the glass holder. The glass sensor
does not recognise the glass when pressing
the glass fill key.
Place a glass to the glass holder. Or change
the setting of the service mode 115.
HE 51
The bowl is above the patient chair when driv-
ing the chair up.
Move the bowl to the rest position.
HE 52
The patient chair can not be driven upwards
because the emergency switch of the side
delivery arm is released.
HE 53
The patient chair can not be driven down-
wards because the emergency switch of the
side delivery arm is released.
HE 54
Glass is inserted in the glass holder.
Remove the glass.
HE 60
Water management system (WMS): Container
almost empty in domestic water supply mode.
Reduce water usage for a moment.
HE 61
WMS: Container almost empty in Clean Water
mode.
Reduce water use and fill up the container as
soon as possible.
HE 62
WMS: The Container switch is turned off.
Switch on the container switch.
HE 63
Cuspidor door is open.
Close the door.
HE 64
Fill the container with disinfectant over the
upper sensor.
HE 65
Fill the container with water over the upper
sensor.
HE 66
Wait 8 hours or switch the unit off for the night.
HE 70
The brushless micromotor is stucked or bro-
ken.
HE 80
Headrest safety operation: length movement
blocked or closing gap has obstacles
Check that nothing is blocking the headrest
movement. The headrest can be driven nor-
mally after the possible obstruction has been
removed.
HE 95
The automatic position of the patient chair has
been programmed to a height which cannot be
reached at the moment because the legrest is
unlocked or the armrests are blocking the
movement.
HE 99
The unit has been switched off without per-
forming the long flushing after water has been
used. The help message will disappear after
the long flushing is performed.
This help message can be disabled, see page
C-38 for details.
End
Flushing is completed. After flushing (short,
long or Waterline Cleaning System): Remove
the syringe from the flushing holder and return
it back. If the syringe is already removed,
return it back to the flushing holder. This
ensures that no water will flow from the
syringes after flushing.
Door
Cuspidor door is open.
Close the door.
Help Code
Meaning
Action or Explanation
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......