Chapter F - ADJUSTMENTS
Planmeca Compact i F-39
WMS OPERATION
Technical Manual
Cup fill+bowl rinse lines and instruments are flushed and hereafter bottle is filled. This is
repeated 4 times.
Countdown timer starts from 4 min. (4 x 1 min.). Timer is stopped during bottle filling.
City Water
Bottled Water
Flushing cycle:
Cup fill + bowl rinse flushing
Similar to city water mode
V1
o
V2
o
V3
o
V4
c
V5
o
Pump
running
Display
timer
Duration
10 s (at same time bowl rinse, V6
open 3 s)
Flushing cycle:
Instrument flushing (each instrument 10 s by
turns, syringe all time)
Similar to City water mode
V1
o
V2
o
V3
o
V4
c
V5
c
Pump
running
Display
timer
Duration
50 s
Flushing cycle:
Bottle filling
Bottle has to be filled manually
V1
o
V2
o
V3
c
V4
c
V5
c
Pump
running
Display
wait
He 65
Duration
Until L1 reached
When cycle is completed
Display: He 36
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......