1.2 Adjusting friction of the OP delivery arm joints
S
C
D
A
B
8 8
8 8
light
E
M
S
rotaf
.eps
To adjust the rotational friction of the console
arm column:
1. Remove the small cover plate from the cus-
pidor cover.
2. Unscrew the two M5x10 screws and
remove the holder plate.
3. Adjust the friction with the adjustment screw
located on the console arm column.
4. Reassemble in reverse order.
Adjust the rotational friction of
the console arm with the 4 mm
allen key. Adjust the two
screws equally to avoid wear-
ing. Tightening the screws
increases the friction.
Adjust the lifting friction of
the console arm with the 4
mm allen key. Tightening
the screw increases the
friction.
Adjust the rotational friction of the instrument
console with the 4 mm allen key. Adjust the
two screws equally to avoid wearing. Tight-
ening the screws increases the friction.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......