Chapter C - SERVICE MODE
Planmeca Compact i C-29
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n 56
Show left (high volume) suction sensor signal for branch 2
(nnn signal)
This is the direct sensor signal from the high volume hose sensor (leftmost) on the suction
hose sensor PCB connected to MCB branch 2 (normally the assistant side suction hose
holder).
The display range is 0 to 255. When a hose is in the holder the display should be in the range
120...255, when no hose is in the holder the signal should be in the range 80... 200.
n 57
Show middle (saliva) suction sensor signal for branch 2
(nnn signal)
This is the direct sensor signal from the saliva suction hose sensor (middle) on the suction
hose sensor PCB connected to MCB branch 2 (normally the assistant side suction hose
holder). See service mode n56 for further information.
n 58
Show right (ejector) suction sensor signal for branch 2 (nnn
signal)
This is the direct sensor signal from the ejector/assistant syringe sensor (right) on the suction
hose sensor PCB connected to MCB branch 2 (normally the assistant side suction hose
holder). See service mode n56 for further information.
n 59
Show AUX input signal for branch 2 (nnn signal)
This is AUX input signal on the suction hose sensor PCB connected to MCB branch 2 (nor-
mally the assistant side suction hose holder). This signal is reserved for future use and shows
always 255.
When the S-key is pressed, special foot control calibration check mode is entered. This mode
is for factory use only. Exit the mode by pressing the S-key again.
2.6 Foot control related service modes
n 60
Show software version in Foot Control CPU (n.nn version)
This is the version of the software embedded in the foot control CPU EPROM. This software
cannot be updated on the field.
n 61
Show foot control switch signals (¦¦ ¦ ¦¦¦¦)
This service mode displays all digital input signals in the foot control. The meaning of the ver-
tical bars (from left to right) is; jumper A, jumper B, AUX input, Chair A, Chair B, Chair C,
Chair D.
A short vertical bar indicates an off state (no signal / no jumper), a long vertical bar indicates
an on state (signal / jumper). Jumpers A and B are optional (not needed) in the Prostyle Com-
pact unit. The AUX in-put is reserved for future use. Chair A,B,C,D signals are convenient for
checking the proper operation of the chair control knob, its magnet and HALL-sensors.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......