Chapter C - SERVICE MODE
Planmeca Compact i C-21
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n.16
Enable factory preset mode (- or S U)
This service mode is used to set the unit to a predefined state only using few keystrokes. This
feature is mainly used at the factory, that is why it is called a “factory preset”. All user pro-
grammable settings and parameters, as well as the most important service mode settings can
be set to a “normal” state. Then the unit can be immediately operated without having first to
do any adjustments or programming. This is useful also on the field if a showroom unit is
installed later for clinical use.
Upon entry this service mode shows a dash (-) in the display. Press the S-key for 2 seconds
(prevents accidental entry) so that the yellow LED lights in the key. The display now shows S
U.
Performing user mode presets
Press the B-key briefly (“below” the U symbol in the display) to perform the user mode presets
(the display shows FP.U during the process). Pressing the B-key sets the factory presets for
the user-controlled settings, which are: period of glass filling, period of bowl rinsing, period of
door open/nurse call, settings of the dentists (chair positions: backrest, height, light), timer
settings, intensity of the operating light and instrument settings (spray, power reduction, chip
blow etc.). After performing the user mode presets the text “turn off” will flash on the display.
Now turn off the unit.
Performing service mode presets
Press the A-key briefly (“below” the S symbol in the display) to perform the service mode pre-
sets (the display shows FP.S during the process). Pressing the A-key sets the factory presets
for all the service mode settings. After performing the service mode presets the text “turn off”
will flash on the display. Now turn off the unit.
n.17
Select display units for pressures and temperature (n
mode = 1, 2, 3, 4)
The unit can show pressures (in the service mode) in either bar, kPa or PSI. Select mode 1 to
show the pressure in bar (factory default), 2 to show the pressures in kPa (kiloPascal), or 3 to
show the pressures in PSI (pounds per square inch). If mode 3 is selected additionally tem-
perature displays (ex. service mode 10) are shown in °Fahrenheit instead of °Celsius. Select
mode 4 to show the pressure in PSI and temperature in °Celsius.
n.18
Reserved for factory use.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......