Chapter D - HELP & ERROR MESSAGES
D-18 Planmeca Compact i
DETAILED DESCRIPTION OF ERROR MESSAGES
Technical Manual
3.7 Foot control related error messages
E 51.n Cable from foot control is loose or connected into wrong
connector on MCB
E 51.1
reserved
E 51.2
Foot control is not responding
The foot control cable is disconnected from the unit, or the cable is open circuited, or the foot
control has failed. Check if the foot control gets power by checking its that its internal yellow
status-LED is blinking. Try to recalibrate the foot control, see page 2.6 “Calibrating foot con-
trol” on page H-15 for details.
E 52
Foot control is of new type, update MAIN software
The recently attached foot control is of a new type, of its software is not compatible with the
MCB software. Please update the MCB software. The new software on the MCB is also still
compatible with the older version(s) of the Prostyle Compact foot controls.
E 53
Foot control data error (cable short circuit)
The data signal is short circuited to ground. Check the condition of the foot control cable and
replace it if necessary. Try with another foot control to determine the location of the problem.
Check also the condition of the internal foot control cable going from the MCB to the foot con-
nector at the rear of the unit base.
E 54.n Foot control calibration error
These errors inform that the foot control calibration failed for some reason. The decimal part
number gives the details:
E 54.1
Foot control EEPROM write failure
E 54.2
Foot control EEPROM read failure
E 54.3
Pedal too far from the sensors, channel A
E 54.4
Pedal too far from the sensors, channel B
E 54.5
Sensor plate is too much tilted to either side, checksum error
E 54.6
Sensor plate is too close to the PCB, channel A
E 54.7
Sensor plate is too close to the PCB, channel B
E 54.8
Foot control re-trigger error.
Calibrate the foot control. If it does not help, check that the capacitance sector
plate on the pedal is not touching the foot control PCB, or is not too close to the
PCB. It is also possible, however very unlikely, that the foot control PCB is
defect.
E 55
reserved
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......