Chapter D - HELP & ERROR MESSAGES
D-2 Planmeca Compact i
HELP MESSAGES
Technical Manual
HE 3
The patient chair cannot be driven down
because the emergency switch is activated.
Check that the space under the seat is free
and nothing is touching the plate at the bot-
tom. The chair can be driven normally after the
possible obstruction has been removed.
HE 4
This left-hand instrument position allows only
a syringe.
The leftmost position in the console is
reserved for syringe only. Please replace this
instrument with a syringe. Place the other
instrument to one of the four right-hand
places.
HE 5
This instrument position does not accept a
syringe.
Please remove the syringe from this position
and replace it with any other kind of instru-
ment. The syringe must be placed to the left-
hand place.
HE 6
The patient chair cannot be driven when an
instrument is in use.
Stop the instrument (release the foot pedal) to
be able to run the chair.
HE 7
The patient chair up/down movement is not
allowed when the cuspidor door is open. Only
the manual backrest movement is allowed.
Close the cuspidor door to be able to run the
chair up/down or to a pre-programmed posi-
tion.
HE 8
The seat height can not be programmed this
high in automatic positions for patient safety
reasons.
Please run the chair slightly down to be able to
program this position into memory.
HE 9
This scaler cannot currently be used with this
unit.
This scaler needs a dedicated electronics that
is not installed, or the scaler electronics
installed is not compatible with this scaler.
HE 10
Please wait while configuring the dental instru-
ment setup.
Please wait a few seconds for the unit to check
your instrument configuration, since it has
changed.
HE 12
The unit’s internal error reporting is disabled.
The unit operates normally but no errors are
reported (exhibition use). Use service mode
15 to turn the error reporting back on. See
page C-20 for details.
HE 13
The instrument cannot run because of a
patient security violation.
Please return the foot control pedal once to
the middle position to start the instrument.
Instruments do not start if the foot control
pedal is already pushed to the right or left
when the instrument is taken out from the
holder.
HE 14
This dental instrument is not recognized by the
unit’s control system (new type of instrument).
To be able to operate this new instrument the
MCB software must be updated. See 5.1 “How
to replace/upgrade the software” on page G-
20 for details.
HE 15
This key has no function at this moment.
Instruments: Please pick up a dental instru-
ment before attempting to alter its settings; or
if the instrument is selected, this function is not
allowed with this instrument.
Other keys: the unit does not have this func-
tion.
Help Code
Meaning
Action or Explanation
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......