Chapter F - ADJUSTMENTS
F-32 Planmeca Compact i
WMS OPERATION
Technical Manual
6 WMS OPERATION
6.1 List of abbreviations
o
= open
c
= closed
Vn
= valve number n, refer to the WMS pneumatical diagram
SWn
= switch number n, refer to the WMS pneumatical diagram
L1
= upper sensor
L2
= lower sensor
6.2 WMS operation, normal use
Continuously
City Water
Bottled Water
SW1
off
SW1
on
SW2
on
SW2
on
V3
o
exception: c when bottle is filled and
instruments not in use
V3
o
V4
c
exception: o for 150 ms, when unit is
turned on (from software version
3.07)
V4
c
exception: o for 150 ms, when unit is
turned on (from software version
3.07)
V7
o
V7
o
Bowl rinse is used, water comes directly from incoming water line.
City Water
Bottled Water
Bowl rinse
Bowl rinse
V1
o
V1
o
V2
c
V2
c
V5
c
V5
c
V6
o
V6
o
Pump
not running
Pump
not running
Display
PCI
Display
PCI
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......