Chapter C - SERVICE MODE
Planmeca Compact i C-37
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
2.11 Other service modes
n.93
Adjustment of the duration of the spray water impulse
This service mode can be used to adjust the duration of the spray water impulse. The use of
this water impulse is explained in section “n 95 Set spray water impulse pressure” below. The
adjustment range is 0 - 20 (correspond to 0.00 - 1.20 sec.), and the factory default is 1 (corre-
sponds to 60 ms).
n.94
Set syringe fibre optic light voltage
This service mode can be used to alter the voltage of the syringe fibre optic light. The range is
0 - 4 V, and the factory default is 3.7 V.
n.95
Set spray water impulse pressure
This service mode can be used to adjust the minimum pressure of the spray water impulse. If
you use only a very small amount of spray water there might be a delay in the spray water
flow. However, you can adjust the water impulse pressure so that the spray water channel is
opened immediately to enable the water to flow without delay. After the opening impulse, the
spray water pressure is taken to the value adjusted when programming the amount of water
for spray. The range is 0 - 30 (compare to spray water settings where 0 = “SH.00” and 30 =
“SH.30”), and the factory default is 15.
n.96
RPM display / Enable low speed function for brush
micromotors (Optional Plus package) (0, 1, 2, 3)
This service mode can be used to enable/disable the RPM display for brush micromotors
(instead of percentage display) and also the low speed function for brush micromotors for
dental units with optional Plus package.
0 = RPM disabled / low speed function for micromotors disabled
1 = RPM disabled / low speed function for micromotors enabled
2 = RPM enabled / low speed function for micromotors disabled
3 = RPM enabled / low speed function for micromotors enabled
The factory default is 0.
n.97
Set suction system cleaning time
This service mode can be used to adjust the cleaning time for the suction system. The availa-
ble range is between 0 sec. (no flushing) and 20 min, and the factory default is 2 min.
n.98
Select how to freeze Intracam image from foot control
This service mode is used to select how Intracam images are frozen using the foot control.
0 = sideways movement of the pedal has no effect
1 = pedal movement activates the door open/nurse call relay (video printer freeze)
2 = pedal movement activates the serial communication connection (2400N81) to Dimaxis
Pushing the pedal to the left will freeze the image, pushing it to the left again will unfreeze the
image.
Pushing the pedal to the right will save the image.
3 = pedal movement activates both the door open/nurse call relay and the Dimaxis serial
communication (1 + 2).
The factory default is 2.
4 - 7 are the same as 0 - 3 but the pedal operation is vertical.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......