Installation
General Notes on Installation
Install the E-727 near the power source so that the power plug can be quickly and easily
disconnected from the mains.
Only use cables and connections that meet local safety regulations.
E-727 models for capacitive sensors:
Electromagnetic signals in the range of the sensor frequency (100 kHz) can interfere with the
sensor signal.
Avoid electromagnetic signals in the range of 100 kHz.
Keep in mind that low-frequency signals may have harmonics in the range of 100 kHz.
If interfering signals in the range of 100 kHz cannot be avoided, take particular care to
ensure suitable shielding and grounding.
Installing the PC Software
The communication between the E-727 and a PC is necessary to configure the E-727 and send
motion commands using the commands of the GCS. Various PC software applications are available
for this purpose.
Performing the Initial Installation
Accessories
PC with a Windows operating system (7, 8, 10) or Linux operating system
Product CD (included in the scope of delivery)
Important information on the procedure for installation on Windows
Before you start installing the PC software on a PC with a Windows operating system (p. 45),
read the following information.