User Manual
E727T0005, valid for E-727
BRO, 2019-06-28
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 73 / 240
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Constant pulse length:
Figure 18: Position Distance" Trigger Mode, constant trigger pulse length
The
Pos. Distance Trig. High Time Definition
parameter has the value 1.
The trigger pulse length is constant. The trigger high time is adjustable using the
Position Distance
Trigger High Time
parameter (ID 0x18000401). The parameter value is given in seconds. If the
parameter value is set to 0 (default setting), the length of a trigger pulse is one servo cycle (50 µs).
Possible values for the
Position Distance Trigger High Time
parameter (ID 0x18000401) depend on
the axis velocity and the TriggerStep value. Pulses can get lost if the pulse length and/or
TriggerStep values are not suitable for the current axis velocity, see figure below for an example.