selected lines are combined via AND bit operation. See "Examples—"TriggerOutAND"
and "TriggerOutOR" Modes" (p. 84).
15= TriggerOutOR; the digital output line <TrigOutID> outputs the signal states of the
output lines selected with TriggerOutMask (<CTOPam> ID 16). The states of the
selected lines are combined via OR bit operation. See "Examples—"TriggerOutAND"
and "TriggerOutOR" Modes" (p. 84).
To select the mode, set <CTOPam> = 3 and <Value> to the code of the mode; default selection is
On Target (2).
Furthermore, it is possible to select the signal polarity for the digital output line (active high /
active low). See "Example—Polarity Setting" (p. 84).
CTO changes the values of the parameters listed below in volatile memory. The current values in
volatile memory can be read with the CTO? command. You can also change these parameters
using SPA (volatile memory) or SEP (non-volatile memory). Furthermore, you can use WPA to copy
the current values from volatile memory to non-volatile memory, where they become the power-
on defaults. When using SPA, SEP or WPA, it is necessary to switch to command level 1 with CCL to
have write access to the parameters. To read the parameter values, you can also query with the
SPA? or SEP? commands.
Parameter ID
Command
Level
Item Type
Concerned
Max. No.
of Items
Data Type Parameter Description
0x18000201
1
Digital Output
Line
3
FLOAT
CTO Trigger Step
(<CTOPam> = 1)
0x18000202
1
Digital Output
Line
3
INT
CTO Axis
(<CTOPam> = 2)
0x18000203
1
Digital Output
Line
3
INT
CTO Trigger Mode
(<CTOPam> = 3)
0x18000205
1
Digital Output
Line
3
FLOAT
CTO Min.Threshold
(<CTOPam> = 5)
0x18000206
1
Digital Output
Line
3
FLOAT
CTO Max.Threshold
(<CTOPam> = 6)
0x18000207
1
Digital Output
Line
3
INT
CTO Polarity
(<CTOPam> = 7)
0x18000208
1
Digital Output
Line
3
FLOAT
CTO Start Threshold
(<CTOPam> = 8)
0x18000209
1
Digital Output
Line
3
FLOAT
CTO Stop Threshold
(<CTOPam> = 9)
0x18000210
1
Digital Output
Line
3
INT
CTO Trigger Out Mask
(<CTOPam> = 16)
The following examples can be reproduced using the command entry facilities of PIMikroMove or
PI Terminal.