The total number of points available for data recording is given by the
Data Recorder Max Points
parameter, ID 0x16000200. The controller allocates these points in equal shares to the available
tables (i.e. to the number of tables given in the TNR? answer). For the E-727, the total number of
points is 262144. If, for example, TNR? replies 8, each table is comprised of 32768 points.
The data recorder sampling period can be read with the RTR? command. The answer gives the
value of the
Data Recorder Table Rate
parameter (ID 0x16000000) whose default value is one
servo cycle. You can cover longer periods by increasing this value. Use the RTR command or change
the parameter value directly in volatile memory.
Data-Recorder Related Commands and Parameters
Command
Description
Notes
DRC
Set Data Recorder
Configuration
Assigns data sources and record options to data
recorder tables in volatile memory (
DRC Data
Source
parameter, ID 0x16000700, and
DRC
Record Option
parameter, ID 0x16000701)
DRC?
Get Data Recorder
Configuration
Reads current data recorder settings
DRL?
Get Number Of Recorded
Points
Reads the number of points comprised by the last
recording.
DRR?
Get Recorded Data Values Note: Reading can take some time, depending on
the number of points.
DRT
Set Data Recorder Trigger
Source
Determines how recording is to be triggered.
Settings will be lost on controller power down or
reboot.
DRT?
Get Data Recorder Trigger
Source
Reads current trigger option
HDR?
Get All Data Recorder
Options
Lists available record options, gives information
about additional parameters and commands
concerned with data recording
IMP
Start Impulse and
Response Measurement
Triggers recording
RTR
Set Record Table Rate
Changes the data recorder table rate in volatile
memory (
Data Recorder Table Rate
parameter, ID
0x16000000)
RTR?
Get Record Table Rate
Reads the current setting of the data recorder
table rate (
Data Recorder Table Rate
parameter,
ID 0x16000000)
STE
Start Step and Response
Measurement
Triggers recording
TNR?
Get Number of Record
Tables
Reads the number of available data recorder
tables (
Data Recorder Chan Number
parameter,
ID 0x16000300)