3
M
o
nitoring
Monitoring
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
3-15
©2006-2015 Physio-Control, Inc.
6
SPO2: UNKNOWN SENSOR
message appears.
A sensor is connected to the
device that is not compatible
with the Masimo Sp0
2
module.
• Refer to
for
sensor compatibility.
7
SPO2: SEARCHING FOR
PULSE
message appears.
A sensor is connected to the
patient and is searching for a
pulse.
• Wait for completion.
8
SPO2: LOW PERFUSION
message appears.
Patient has a weak pulse.
• Change sensor location.
Table 3-3
Troubleshooting Tips for SpO
2
(Continued)
Observation
Possible Cause
Corrective Action