Data Management
6-14
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
TROUBLESHOOTING TIPS FOR DATA TRANSMISSION
Table 6-4
Troubleshooting Tips for Data Transmission
Observation
Possible Cause
Corrective Action
1
SEND DATA
option does
not appear in Options/
Archives screen.
CodeManagement Module not
properly connected to
defibrillator.
• Contact qualified service
personnel.
2
TRANSMISSION FAILED
message appears.
No wireless connection to
LIFENET System.
• Verify Wireless switch on
side of CodeManagement
Module is in the
ON
position.
CodeManagement Module,"
page 2-16
• Contact your facility’s
LIFENET System
administrator to verify the
receiving computer is
operational.