LIFEPAK 20e Defibrillator/Monitor Operating Instructions
7-11
©2006-2015 Physio-Control, Inc.
11
Date printed on report
is incorrect.
Date is incorrectly set.
• Change the date setting. Refer to
12
Displayed messages
are faint or flicker.
Low battery power.
Out of temperature range.
• Connect to AC power immediately.
13
Low speaker volume.
Moisture in speaker grill
holes.
• Wipe moisture from speaker grill and
allow device to dry.
14
MAINTENANCE DUE
message appears.
Maintenance prompt is set
to display at a selected
interval in service mode.
• Continue to use device if needed.
• Contact service personnel to reset or
turn off the maintenance prompt.
• Contact Physio-Control Technical
Support for instructions on how to
reset or turn off this prompt.
15
SELF TEST DID NOT
COMPLETE
message
appears.
Test plug not connected to
QUIK-COMBO therapy
cable during daily auto test.
• Connect test plug to QUIK-COMBO
therapy cable.
• Perform manual user test.
Standard paddles not
seated in paddle wells
during daily auto test.
• Make sure standard paddles are
securely seated in the paddle wells
for daily auto test.
Defective therapy cable or
a problem with the
defibrillator.
• Contact qualified service personnel.
16
SELF TEST FAILED
message appears.
Device self-test circuitry
detects service condition
during daily auto test.
• Use defibrillator or pacemaker if
needed in an emergency.
• Report occurrence of self-test failure
to qualified service personnel.
17
USER TEST DID NOT
COMPLETE
message
appears.
Test plug not connected to
QUIK-COMBO therapy
cable during user test.
• Connect test plug to QUIK-COMBO
therapy cable for user test.
Standard paddles not
seated in paddle wells
during user test.
• Make sure standard paddles are
securely seated in the paddle wells
for user test.
Defective therapy cable or
a problem with the
defibrillator.
• Contact qualified service personnel.
18
USER TEST FAILED
message appears.
Device self-test circuitry
detects service condition
during user test.
• Use defibrillator or pacemaker if
needed in an emergency.
• Report occurrence of user-test
failure to qualified service personnel.
19
Printout annotated
79 J, but > 79 J was
selected.
Paddles discharged when
seated in the paddle wells
or shorted together.
Defibrillation and Synchronized
Cardioversion Check on Battery
Power," page 7-6
.
• Perform User Test.
20
Defibrillator does not
charge to 200 J within
10 seconds while on
battery power.
Low battery condition.
Defective battery.
• Connect the defibrillator to AC
power.
• Contact qualified service personnel.
Table 7-2
General Troubleshooting Tips (Continued)
Observation
Possible Cause
Corrective Action
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......