Therapy
4-8
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
You will see and hear
STAND CLEAR,
PUSH TO SHOCK
(
) followed by a
“shock ready” tone. The shock LED
flashes.
Clear everyone away from the patient,
bed, or any equipment connected to the
patient.
Press the
button to discharge the
AED.
Note:
If you do not press the
button within 60 seconds, the AED
disarms the shock button, and the
DISARMING
message appears.
When the
button is pressed you will
see the message
ENERGY DELIVERED
indicating energy transfer was
completed.
When energy transfer is complete the
shock counter increases by 1. This will
continue to increase incrementally with
every energy transfer.
After a shock is delivered you will see
and hear
START CPR
. A countdown
timer (min:sec format) continues for the
duration specified in the
CPR TIME 1
setup option.
Note:
If the CPR metronome is set
to be on during CPR times, you will
hear audible “tocks” and ventilation
prompts or tones.
When the CPR countdown time ends,
you will see and hear
PUSH ANALYZE
.
This message stays on the screen and
the voice prompt will repeat every 20
seconds until you press the
ANALYZE
button.
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......