SpO2 Clinical Validation Summaries
F-2
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
NELLCOR CLINICAL VALIDATION DATA
Nellcor conducted clinical studies and tests to assess the accuracy of the SpO
2
measurement
function of the Nellcor SpO
2
sensors. The following paragraphs provide a summary of that data.
Test Methods for Accuracy
SpO
2
accuracy specifications for Nellcor sensors are based on controlled hypoxia studies with
healthy nonsmoking adult volunteers over the specified saturation SpO
2
range. Pulse oximeter
SpO
2
readings were compared to SaO
2
values of drawn blood samples measured by
hemoximetry.
Subjects used to validate SpO
2
measurement accuracies were healthy and recruited from the
local population. The study group comprised both men and women; subjects spanned a range of
skin pigmentations and ranged in age from 18-50 years old. When sensors are used on neonatal
subjects as recommended, the specified accuracy is decreased by ±1%, as compared to adult
usage, to account for the theoretical effect on oximeter measurements of fetal hemoglobin in
neonatal blood.
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......