Setup Options
8-6
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
To activate, select
ENERGY PROTOCOL
in the Paddle Default menu. Auto energy sequences
are disabled if you press the
ENERGY SELECT
control or change to or from AED mode during
use.
Table 8-4
Manual Mode Energy Protocol Setup Menu
Menu Item
Help Message
Options
PRESET
PROTOCOL
Select preset energy
protocol
FULL RANGE
,
PEDIATRIC
.
ENERGY 1
Select energy level for
shock 1
Full range:
100, 125, 150, 175
,
200
,
225,
250, 275, 300, 325, 360
.
Pediatric:
2, 3, 4
,
5
,
6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100
.
ENERGY 2*
Select energy level for
shock 2
Full range:
100, 125, 150, 175, 200, 225,
250, 275
,
300
,
325, 360
.
Pediatric:
2
,
3
,
4
,
5
,
6
,
7
,
8
,
9
,
10
,
15
,
20
,
30
,
50
,
70
,
100
.
ENERGY 3*
Select energy level for
shock 3
Full range:
100, 125, 150, 175, 200, 225,
250, 275, 300, 325
,
360
.
Pediatric:
2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30,
50, 70, 100, 125, 150
.
*
ENERGY 2
cannot be less than
ENERGY 1
.
ENERGY 3
cannot be less than
ENERGY 2
.
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......