Appendix A
Specifications and Performance Characteristics
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
A-3
©2006-2015 Physio-Control, Inc.
MONITOR
ECG
ECG is monitored via several cable arrangements. A 3-wire
cable is used for 3-lead ECG monitoring. A 5-wire cable is
used for 3-lead ECG plus AVR, AVL, AVF, and C. Standard
paddles or therapy electrodes (QUIK-COMBO pacing/
defibrillation/ECG electrodes or FAST-PATCH disposable
defibrillation/ECG electrodes) are used for Paddles lead
monitoring. Compatible with LIFEPAK 12 ECG and Therapy
Cables.
Lead Selection
Leads I, II, III (3-wire ECG cable)
Leads I, II, III, AVR, AVL, AVF, and C acquired
simultaneously (5-wire ECG cable)
ECG Size
4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV
Heart Rate Display
20–300 bpm digital display
Out of range indication: Display symbol "---"
Heart symbol flashes for each QRS detection
Recovery Time after
Defibrillation
10 seconds
Continuous Patient
Surveillance System
(CPSS)
In AED mode, while Shock Advisory System is not active,
CPSS monitors the patient, via QUIK-COMBO paddles or
Lead II ECG, for potentially shockable rhythms.
Voice Prompts
Used for selected warnings and alarms (configurable on/off).
Analog ECG Output
1 V/mV x 1.0 gain <35 msec delay. No internal pacemaker
pulse enhancement or detection markers included.
Common Mode
Rejection
90 dB at 50/60 Hz
Pacemaker Pulse
Rejection
Rejects pacemaker pulses having amplitudes from ±2 mV to
±700 mV and pulse widths from 0.1 ms to 2.0 ms with and
without overshoot. Pacemaker pulse overshoot is defined as
2.5% to 25% of the pacemaker pulse amplitude not to exceed
2 mV. Refer to IEC 60601-2-27.
Note:
Does not apply when internal paddles are connected.
SpO
2
Masimo
®
sensors
Saturation Range
1 to 100%
Saturation accuracy is specified for range 70–100% (0–69% unspecified)
Adults/Pediatrics
Accuracy (RMS)*
±2% (during no motion conditions - Masimo)
±3% (during no motion conditions - Nellcor)
±3% (during motion conditions - Masimo)
Neonatal Accuracy
(RMS)*
±3% (during no motion conditions - Masimo)
±4% (during no motion conditions - Nellcor)
±3% (during motion conditions - Masimo)
*Because the above SpO
2
measurements are statistically distributed, only about
two-thirds of the measurements can be expected to fall within the accuracies
specified above.
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......