Specifications and Performance Characteristics
A-4
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Dynamic Signal
Strength Bar Graph
Pulse Tone at the
Onset of the Pleth
Waveform
SpO
2
Averaging Time
User selectable 4, 8, 12 or 16 seconds
SpO
2
Data Update
Period
1 second
SpO
2
Alarm Condition
Delay
21 seconds (maximum time from physiological change to
detection by device, with default 8 second averaging
selected)
SpO
2
Alarm Signal
Generation Delay
1 second (time for device to generate alarm after alarm
condition is detected)
SpO
2
Measurement
Functional SpO
2
values are displayed and stored
Pulse Rate Range
25 to 240 pulses per minute
Pulse Rate Accuracy
Adults/Pediatrics/
Neonates
±3% (during no motion conditions)
±5% (during motion conditions)
SpO
2
Waveform with
Autogain Control
SpO
2
Measurement Wavelengths
Note:
Information about wavelength range can be useful to clinicians, for example,
when performing photodynamic therapy.
Masimo
Red: 660 nanometers
Infrared: 905 nanometers
Nellcor
Red: 660 nanometers
Infrared: 900 nanometers
SpO
2
Optical Power
Masimo
Maximum optical output power = 15 mW
Nellcor
Maximum optical output power = 15 mW
Alarms
Quick Set
Activates alarms for all parameters
VF/VT Alarm
Activates continuous CPSS monitoring in Manual Mode
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......