Screen Messages
C-2
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
CONNECT SYNC
CABLE TO REMOTE
MONITOR
Remote sync is selected and the device is not connected to the
remote monitor.
CONNECT TO AC
POWER
Remote sync is selected and the device is not connected to AC power.
CONNECT TO TEST
PLUG
Test plug not connected to QUIK-COMBO therapy cable or standard
paddles not seated in paddle wells during user test.
CPR: ADULT-AIRWAY
X:Y
An option for CPR metronome. The patient is an adult for whom an
advanced airway has been established. The specified C:V ratio will be
used.
CPR: ADULT-NO
AIRWAY X:Y
An option for CPR metronome. The patient is an adult for whom an
advanced airway has not been established. The specified C:V ratio will
be used.
CPR: YOUTH-AIRWAY
X:Y
An option for CPR metronome. The patient is a youth (younger than
the age of puberty) for whom an advanced airway has been
established. The specified C:V ratio will be used.
CPR: YOUTH-NO
AIRWAY X:Y
An option for CPR metronome. The patient is a youth (younger than
the age of puberty) for whom an advanced airway has not been
established. The specified C:V ratio will be used.
CURRENT FAULT
The comparison between delivered and selected pacing current is out
of tolerance.
DEMAND
The pacemaker is in the demand mode.
DISARMING...
The decision was made to remove the energy charge.
ECG CABLE OFF
The ECG cable was removed during printing.
ECG LEADS OFF
Multiple ECG electrodes were disconnected either when the device
was turned on or during monitoring.
ENERGY DELIVERED
Energy transfer was completed.
ENERGY NOT
DELIVERED
An open air discharge is detected with standard paddles. Usually this
is because the electrodes are not in contact with a patient or test load
when the shock is initiated.
ENERGY SELECT/XXX
J
The front panel or standard paddles
ENERGY SELECT
button was
pressed.
IF NO PULSE, PUSH
ANALYZE
This message appears following a CPR interval (if activated in setup
configuration).
IF NO PULSE, START
CPR
Initiate CPR if no pulse and continue with CPR until completion tone.
IF YOU WITNESSED
THE ARREST, PUSH
ANALYZE
Initial CPR message following
START CPR
prompt to remind user to
deliver a shock immediately, if the user witnessed the arrest.
Table C-1
Summary of Screen Messages (Continued)
Message
Description
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......