LIFEPAK 20e Defibrillator/Monitor Operating Instructions
4-9
©2006-2015 Physio-Control, Inc.
No Shock Advised
Proper Connection
If the AED detects a nonshockable
rhythm, you will see and hear
NO
SHOCK ADVISED
. The AED will not
charge, and a shock can not be
delivered.
After a
NO SHOCK ADVISED
prompt you
will see and hear
START CPR
. A
countdown timer (min:sec format)
continues for the duration specified in
the
CPR TIME 2
setup option. (Refer to
Note:
If the CPR metronome is set
to be on during CPR times, you will
hear audible “tocks” and ventilation
prompts or tones.
When the CPR countdown time ends,
you will see and hear
PUSH ANALYZE
.
This message stays on the screen and
the voice prompt will repeat every 20
seconds until you press the
ANALYZE
button.
If therapy electrodes are not connected
to the therapy cable or not placed on the
patient’s chest, the
CONNECT
ELECTRODES
message and voice
prompt occur until the patient is
connected to the AED.
If the therapy cable is not connected to
the defibrillator, you will see the
CONNECT CABLE
message until the
cable is connected.
If the test plug is connected to the
therapy cable when analysis is initiated,
the
REMOVE TEST PLUG
message and
voice prompt occurs.
Содержание LIFEPAK 20
Страница 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Страница 3: ......
Страница 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Страница 15: ......
Страница 49: ......
Страница 111: ......
Страница 155: ......
Страница 171: ......
Страница 181: ......
Страница 183: ......
Страница 189: ......
Страница 191: ......
Страница 195: ......
Страница 199: ......
Страница 201: ......
Страница 205: ......
Страница 209: ......
Страница 211: ......
Страница 213: ......
Страница 215: ......
Страница 221: ......
Страница 226: ......