Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
4 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
Work Instruction Icon Definitions:
Work
Content
1
1
st
Person
Verify
Torque
!
Caution
Shock
Hazard
E
N
S
W
Orient
Clean
Work Content:
The value add activity being done at the operation. Create a text box in this color to highlight the parts being
added.
1st Person Verify: Checking a person
’s own work. This can be added as part of a pair of verifications driven from risk
assessment or can be called out by itself based upon some business need. Good verification practice
involves multiple senses (usually touch and look).
Torque:
Use with any fastener that has a specific torque specification. Call out the specific torque requirement in
association with this symbol (measure and units). A fastener can be screws, bolts, nuts, or another
mechanical fastening device.
Caution:
When a specific operation requires extra attention. It can be something that is challenging to perform, has a
higher likelihood of being performed incorrectly, or could present some level of risk to the person performing
the process.
Shock Hazard:
When a person is exposed to something that contains electrical energy. Most likely during disassembly after
a unit has been energized.
Orient:
When the orientation of the part matters and it could be assembled multiple ways. Making sure that labeling is
readable is the most common example.
Clean:
For cosmetics or for adhesive preparation. The default cleaning material is chem-wipes. Call out if any
additional materials are to be used (such as alcohol) and the purpose of the cleaning.
1. Verify the Device Serial Number
Verify that device’s serial number label matches the serial number listed in the upgrade kit documentation.
2. Initial Power-Up Check
Install a fully charged battery onto the LIFEPAK 15 Monitor/Defibrillator. Press the
ON
key. Verify that the
device powers up with no service indicator light. If the service indicator light is illuminated, do not proceed
with this upgrade, and repair per the appropriate LIFEPAK 15 Monitor/Defibrillator repair process. If the
service indicator light is not illuminated, power off the device, remove the battery and continue with upgrade.
3. Downloading Device Data (Pre-Service Report Download)
FOR INTERNATIONAL, Use your normal service reporting process to document the service activity.
Open a service report and download the device data to the service report before proceeding with disassembly
of the device.
TO
R
Q
U
E
TO
R
Q
U
E
TO
R
Q
U
E
XX.X In-Lb