4.
Disassembling the Case (continued)
4.8
Disconnect the P2 system/interface flex cable
connector (W04) from the system PCB in the
rear case.
4.9
Disconnect the P23 therapy ribbon cable
connector (W11) in the front case from the
therapy PCB.
4.10 Disconnect the J24 therapy cable connector
(W11) from the P24 transfer relay connector
(A13).
4.11 Separate the front and rear cases halves.
Disconnect
P30
– System/
Interface Flex
Disconnect P2
–
System/Interface
Flex (W04). Lift
connector tab to
disconnect cable.
Disconnect
J24 - Therapy
Cable (W11)
Disconnect
P23 - Therapy
Ribbon Cable
(W11)