Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
25 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
15. Installing the System (A01)/Therapy (A04) PCB
Assembly
To install the system/therapy PCB:
15.1 Line up the system/therapy assembly with rear
case.
15.2 Connect the W14 - USB flex cable (3206966)
to J4 of the system PCB.
15.3 Connect the W01 - power/system cable to J1 of
system PCB.
15.4 Connect the W02 - power/therapy cable from
the power PCB P20 (3009726) to J20 of the
therapy PCB.
15.5 Connect the 5-pin connector from the A13 -
transfer relay P18 to J18 of the therapy PCB.
15.6 Connect the 5-pin connector from the A15 -
energy capacitor to J19 of the therapy PCB.
Connect J4 USB
Flex (W14)
Connect J1
Power/System
Cable (W01)
Connect J20
Power/Therapy
Cable (W02)
Connect J18
Transfer Relay
Cable (A13) and
J19 Energy Cap
Cable (A15)