Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
12 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
6.
Removing the OEM PCB (A06)
6.1
Disconnect Power Therapy Cable (W02) from
J8 on the power PCB.
6.2
Remove the OEM shield (3208298).
6.3
Remove the nylon standoff (3010569-02) from
the OEM PCB.
6.4
If present, disconnect the SpO2/OEM cable
(W21).
6.5
If present, remove the NIBP connector retainer
clip (3010805-000) and disconnect the
NIBP/OEM cable (W27).
6.6
Remove the four mounting screws. Discard the
screws.
6.7
Lift the right edge of the OEM PCB upward
slightly, and then slide the PCB to the right
(toward the capacitor) to disengage the direct
connection between the OEM PCB at J25 and
the power PCB at J16.
6.8
Lift the OEM board from the rear case.
Disconnect J8
Power/ Therapy
Cable (W02)
Remove
OEM Shield
Disconnect J28 NIBP/
OEM Cable (W27)
Remove nylon
standoff
Disconnect J26 SpO2/
OEM Cable (W21)
Disconnect J27 CO2/
OEM Cable (W27)
Remove Screws
(4 places)