Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
16 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
10. Installing the CO2 Inlet Connector Cable (W28)
10.1 Set the CO2 connector adapter (3012119) in
place on the face of the bezel.
10.2 Place the new CO2 connector retainer
(3012121) and new seal (3007997-01) in place
on the underside of the bezel and tighten
together with four new screws (202253-732),
torque to 32 in-oz using a P0 bit.
CAUTION
POSSIBLE MOISTURE LEAKAGE
- When installing
the W28 cable, apply 32 in-oz of torque when
installing the screw retaining the W28 cable to help
prevent ingress of fluids.
10.3 Slide the CO2 inlet connector cable (W28)
(3012140-008) into the new CO2 connector
retainer.
CO2 Connector
Adapter 3012119
Screw
202253-732
(4 places)
CO2 Connector
Retainer 3012121
CO2 Seal
3007997
CO2 Inlet Connector
Location
FRS Assembly
3012140-008
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
32 In-Oz