11. Installing the Parameter Bezel
Observe the following:
Clean away any adhesive residue on the
parameter bezel with isopropyl alcohol.
CAUTION
POSSIBLE MOISTURE LEAKAGE
- Visually inspect
the mating surfaces between the parameter bezel and
the rear case before and after fastening to ensure that
they are even.
11.1 Place the parameter bezel above the rear case
to make the cable connections.
11.2 If present, route NIBP tube through opening in
power PCB bracket next to the J9 connector,
power PCB.
11.3 If present, connect the SpO2 connector cable
(W22) to SpO2 PCB (A16) at JP1.
11.4 Route W28 CO2 inlet tube and cable around
end of Power PCB and above SpO2 cable.
11.5 Reinstall the parameter bezel onto rear case
with two new screws (202253-763), torque to
6.8 in-lb.
CAUTION
Ensure the NIBP hose isn't pinched by the ECG
Brackets or rear case standoff. NIBP hose movement
should be unrestricted in bezel cavity.
Screw
202253-763
(2 places)
Route CO2 Inlet Cable
and Tubing around
edge of Power PCB
and above SpO2 cable
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
6.8 In-Lb
NIBP hose
routing