Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
30 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
16. Reassembling the Case (continued)
16.6 Install 14 new screws (201407-069), torque to
10.0 in-lb using a P2 bit.
16.7 Reinstall the right corner bumper (3207307-
001) with 4 screws (3207361-375), torque to
10.0 in-lb using Torx T-15 bit.
16.8 Reinstall the left corner bumper (3206968-000)
with 4 screws (3207361-375), torque to 10.0 in-
lb using Torx T-15 bit.
NOTE:
If new bumpers and feet are required,
use Guard & Feet Repair Kit (MIN 3305431-
019).
Screws
201407-069
(14 places)
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
10.0 In-Lb
Screws
3207361-375