Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
3 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
Materials Required:
Part Number
Description
Qty
3207813-XXX
Or
3305642-XXX
(V1) LABEL SET-(configured per kit)
(V2) LABEL SET-(configured per kit)
1
3012121-002
RETAINER-CO2 CONN
1
3012140-006
MODULE, CO2, MINI
1
202253-760
SCREW-M,PH,NYLOK,CS,4-40,.250L
4
3012397-01
CABLE ASSY-CO2, ADAPTER
1
3007997-01
SEAL-CO2 CONNECTOR
1
3012119-02
ADAPTER-CO2 CONNECTOR
1
3012140-008
FRS ASSEMBLY-CO2, MINI-MODULE, LP15
1
3206998-002
CABLE ASSY-CO2 MODULE, OEM PCB
1
3012140-002
EXHAUST TUBING-CO2
1
202253-732
SCREW-MACH,NYLOK,PNH,PHH,2-56X0.500,CS,ZN
4
3207337-312
SCREW-MACH,PNH,PHH,NYLOCK,4-40,0.312L,WSHR,CS,ZN
4
201407-069
SCREW,SELF-SEAL,SELF-LOCK,6-32 X 0.375
14
202253-760
SCREW-M,PH,NYLOK,CS,4-40,.250L
8
202253-761
SCREW-M,CS,Z,PH, NYLOCK, 4-40 X .312L 12
13
202253-763
SCREW,MACHINE,PANHEAD,NYLOK,4-40 X .437
2
3207361-375
SCREW-CAP,SCH,REC 15IP TORX PLUS,6-32,0.375L,SS
8
Warnings and Cautions:
LETHAL VOLTAGES MAY BE PRESENT INSIDE THIS DEVICE. THE STORAGE CAPACITOR MUST BE
DISCHARGED PRIOR TO PERFORMING ANY WORK WITHIN THE REAR CASE ASSEMBLY.
FAILURE TO DISCHARGE THE STORAGE CAPACITOR MAY RESULT IN INJURY OR DEATH.
General Requirements:
Unless specifically noted, all removed fasteners are to be discarded and new fasteners used for reassembly.
All removed parts with adhesive backing are to be discarded and new parts used for reassembly.
Unless specifically noted, all removed parts, except fasteners and adhesive backed parts, are to be kept with the device being
upgraded and re-used to reassemble the device.
Inspect all removed components for damage during disassembly and replace all damaged parts with new parts.