Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
29 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
16. Reassembling the Case
To reassemble the case halves:
16.1 Position the front case over the rear case.
16.2 Connect the P23 therapy ribbon cable
connector (W11) (3207044) from the front case
to J23 of therapy PCB.
16.3 Connect the J24 6-pin block therapy cable
connector (W11) to P24 transfer relay
connector (A13).
16.4 Carefully connect the P2 system/interface flex
cable connector (W04) (3206991) to the
system PCB in the rear case.
16.5 Ensure the case seal is in place in the groove
along outside edge of front case. Fold the front
and rear case halves together. Route the
therapy cable wires away from the edge of the
case to avoid pinching.
CAUTION
POSSIBLE COMPONENT DAMAGE
- Ensure the
therapy cable wires are routed away from the edge of
case to avoid wire pinching.
POSSIBLE MOISTURE LEAKAGE
- Visually inspect
the mating surfaces between the front and the rear
case halves before and after fastening them together
to ensure that they are even.
PREVENT VIBRATION DAMAGE
- In order to meet
vibration specifications (for example, prevent
loosening of case screws), use new screws (201407-
069) when assembling the case and torque to
specified value.
Connect J24
Therapy
cable (W11)
Connect J23
Therapy Ribbon
cable (W11)
1
Connect P2
System/ Interface
Flex (W04)