Troubleshooting Guide
54
Chapter 5 - Troubleshooting
3
NiBP Pressure Violation Error.
• The pressure on the cuff
could not be removed by
normal means.
• A fault has been detected in
the NiBP system that could
not be handled by releasing
pressure by normal means.
• NiBP system has released
pressure using safety valve.
1. Cycle power.
2. Check for blocked tubing in
the pneumatic system.
3. Replace Main PCB.
Note: This error may occur
during a blood
pressure measurement
in STAT mode, if the leg
cuff is used.
4
The measured value of the 3.3V
power supply is low.
1. Check power supply.
2. Replace power supply
assembly.
5
The measured value of the 3.3V
power supply is high.
1. Check power supply.
2. Replace power supply
assembly.
6
The measured value of the 12V
power supply is low.
1. Check power supply.
2. Replace power supply
assembly.
7
The measured value of the 12V
power supply is high.
1. Check power supply.
2. Replace power supply
assembly.
8
The measured value of the 5V
power supply is low.
1. Check power supply.
2. Replace power supply
assembly.
9
The measured value of the 5V
power supply is high.
1. Check power supply.
2. Replace Main PCB.
A
The measured value of the
isolated reference supply on the
front end is low.
1. Check power supply.
2. Replace Main PCB.
B
The measured value of the
isolated reference supply on the
front end is high.
1. Check power supply.
2. Replace Main PCB.
C
The measured value of the
isolated reference supply on the
front end is high.
1. Check power supply.
2. Replace Main PCB.
64
The SpO
2
module is sending an
error message to the host CPU.
1. Ensure MP-205 module is
properly connected.
2. Replace MP-205 module.
3. If problem persists, replace
Main PCB.
Hex Code
Explanation
Recommended Action
Содержание M3921A
Страница 12: ...xii Contents ...
Страница 16: ...Related Documents 4 Introduction ...
Страница 48: ...Safety Tests 36 Performance Verification ...
Страница 72: ...Troubleshooting Guide 60 Chapter 5 Troubleshooting ...
Страница 99: ...Disassembly Procedures A3 Chapter 6 Disassembly Guide 87 NiBP pump assembly Main PCB Main ribbon cable ...
Страница 100: ...Disassembly Procedures A3 88 Chapter 6 Disassembly Guide ...
Страница 102: ...Top Level Assembly 90 Chapter 7 Spare Parts Top Level Assembly Figure 1 A1 Top Level Assembly Drawing 1 of 2 ...
Страница 103: ...Top Level Assembly Chapter 7 Spare Parts 91 Figure 2 A1 Top Level Assembly Drawing 2 of 2 ...
Страница 105: ...Top Level Assembly Chapter 7 Spare Parts 93 Figure 3 A3 Top Level Assembly Drawing ...
Страница 107: ...Front Case Assembly Chapter 7 Spare Parts 95 Front Case Assembly Figure 4 A1 Front Case Assembly Drawing 1 of 3 ...
Страница 108: ...Front Case Assembly 96 Chapter 7 Spare Parts Figure 5 A1 Front Case Assembly Drawing 2 of 3 ...
Страница 110: ...Front Case Assembly 98 Chapter 7 Spare Parts Figure 7 A3 Front Case Assembly Drawing 1 of 2 ...
Страница 111: ...Front Case Assembly Chapter 7 Spare Parts 99 Figure 8 A3 Front Case Assembly Drawing 2 of 2 ...
Страница 114: ...Rear Case Assembly 102 Chapter 7 Spare Parts Figure 9 A3 Rear Case Assembly Drawing 1 of 3 ...
Страница 115: ...Rear Case Assembly Chapter 7 Spare Parts 103 Figure 10 A3 Rear Case Assembly Drawing 2 of 3 ...
Страница 116: ...Rear Case Assembly 104 Chapter 7 Spare Parts Figure 11 A3 Rear Case Assembly Drawing 3 of 3 ...
Страница 124: ...Repacking In A Different Carton 112 Chapter 8 Packing For Shipment ...
Страница 132: ...Theory of Operation and System Architecture 120 Chapter 10 Training Program Block Diagram A1 Figure 1 A1 Block Diagram ...
Страница 144: ...Preventative Maintenance 132 Chapter 10 Training Program ...
Страница 150: ......