Troubleshooting Guide
Chapter 5 - Troubleshooting
51
Error Codes
When the monitor detects an error condition, the monitor shows an error code
on the display screen. If such an error occurs during monitoring operation, an
audible alarm tone also sounds. Press the
ALARM SILENCE
button
to
terminate the audible alarm tone.
When an error code appears on the display, a number in hexadecimal
representation indicates the nature of the error. Additionally, Diagnostic Mode
can be used to gain access to an error code record, stored in nonvolatile
memory, of the last 10 error codes encountered by the monitor. See Chapter 4
for further details on Diagnostic Mode.
Each error code corresponds to a particular problem in the monitor.
Recommended actions to take when an error code is encountered are listed in
the sections that follow.
As an aid to troubleshooting, the A1 provides the capability for technicians to
print out a copy of the error log.
Generating an Error Log Printout (A1 only)
1. Connect the recorder to the monitor, and its power supply to an
appropriate source. Refer to the operator’s manual.
2. Use the
SETUP
button
and displayed menu to verify that the
Recorder mode is the selected option for the Communications Selection
item. (It is the factory-set default value.)
3. Turn monitor power OFF.
2. The A3 turns on,
then shuts off and
sounds an alarm
and no error code is
displayed.
1. Press Alarm Silence switch to terminate audible
alarm. Ensure AC power cord is connected and green
BATTERY CHARGING/AC SOURCE
indicator is lit,
or ensure DC source is connected and green
BATTERY CHARGING/DC SOURCE
is lit. If monitor
operates successfully, battery can be discharged.
2. Recharge the battery as directed in “Battery Charge”
on page 11. If the battery fails to hold a charge,
replace the battery.
3. If the problem persists, replace the main PCB.
4. If the problem persists, replace the power supply
assembly.
Condition
Recommended Action
Содержание M3921A
Страница 12: ...xii Contents ...
Страница 16: ...Related Documents 4 Introduction ...
Страница 48: ...Safety Tests 36 Performance Verification ...
Страница 72: ...Troubleshooting Guide 60 Chapter 5 Troubleshooting ...
Страница 99: ...Disassembly Procedures A3 Chapter 6 Disassembly Guide 87 NiBP pump assembly Main PCB Main ribbon cable ...
Страница 100: ...Disassembly Procedures A3 88 Chapter 6 Disassembly Guide ...
Страница 102: ...Top Level Assembly 90 Chapter 7 Spare Parts Top Level Assembly Figure 1 A1 Top Level Assembly Drawing 1 of 2 ...
Страница 103: ...Top Level Assembly Chapter 7 Spare Parts 91 Figure 2 A1 Top Level Assembly Drawing 2 of 2 ...
Страница 105: ...Top Level Assembly Chapter 7 Spare Parts 93 Figure 3 A3 Top Level Assembly Drawing ...
Страница 107: ...Front Case Assembly Chapter 7 Spare Parts 95 Front Case Assembly Figure 4 A1 Front Case Assembly Drawing 1 of 3 ...
Страница 108: ...Front Case Assembly 96 Chapter 7 Spare Parts Figure 5 A1 Front Case Assembly Drawing 2 of 3 ...
Страница 110: ...Front Case Assembly 98 Chapter 7 Spare Parts Figure 7 A3 Front Case Assembly Drawing 1 of 2 ...
Страница 111: ...Front Case Assembly Chapter 7 Spare Parts 99 Figure 8 A3 Front Case Assembly Drawing 2 of 2 ...
Страница 114: ...Rear Case Assembly 102 Chapter 7 Spare Parts Figure 9 A3 Rear Case Assembly Drawing 1 of 3 ...
Страница 115: ...Rear Case Assembly Chapter 7 Spare Parts 103 Figure 10 A3 Rear Case Assembly Drawing 2 of 3 ...
Страница 116: ...Rear Case Assembly 104 Chapter 7 Spare Parts Figure 11 A3 Rear Case Assembly Drawing 3 of 3 ...
Страница 124: ...Repacking In A Different Carton 112 Chapter 8 Packing For Shipment ...
Страница 132: ...Theory of Operation and System Architecture 120 Chapter 10 Training Program Block Diagram A1 Figure 1 A1 Block Diagram ...
Страница 144: ...Preventative Maintenance 132 Chapter 10 Training Program ...
Страница 150: ......