Diagnostic Mode
42
Power-up Defaults Menu And Diagnostic Mode
operational. This value can be reset to zero, for
instance at the time when a technician installs a
new recorder.
•
Battery Deep Discharges
:
Displays the number of deep-discharge cycles seen
by the battery. The monitor records a deep
discharge cycle when the battery volt-age reaches
5.6 volts, the voltage at which a
Low Battery
alarm is issued. This value can be reset to zero, for
instance at the time when a technician installs a
new battery.
•
System Software Version
: Displays the revision level of the system software.
The revision level is also momentarily shown on
the LCD as part of the Copyright screen. This value
cannot be changed by the user.
•
SpO
2
Software Version
:
Displays the revision level of the software of the
MP-205 SpO
2
module. This value cannot be
changed by the user.
When in the System Information screen, the wheel can be rotated to select any of the
changeable items. If one of those items is selected, a press of the wheel causes a pop-
up menu to appear. The first item in the pop-up reads
Make no change
; the
second item in the pop-up reads
Reset to zero
. Exiting the screen is
accomplished in the normal manner, by selecting
Return
.
System A/D Values
The System A/D screen displays the current value of each analog-to-digital (A/D)
channel in volts. Some of the channels are for AC-coupled signals (such as ECG
input), so the numbers on the screen are constantly changing when an input signal is
present. These AC-coupled values are shown to give an indication as to whether basic
functionality of the channel is present, but no significance can be derived from the
values of the numbers displayed. However, others of the A/D channels read DC
voltages, (for example, power supply voltages and battery voltage) and those voltage
values directly provide useful diagnostic information.
The Primary and Secondary Status messages from the SpO
2
module are displayed and
updated at the rate of about once per second. Presence of the correct SpO
2
message
indicates that, at a basic level, communication between the SpO
2
module and the main
monitor processor is working correctly. None of the displayed values can be changed
or reset in this screen.When in the System A/D screen, the
Return
item is always
highlighted; a press of the wheel returns the user to the Diagnostic Menu. Rotating the
wheel while in the System A/D screen has no effect. The A/D channel designators are
shown below.
Содержание M3921A
Страница 12: ...xii Contents ...
Страница 16: ...Related Documents 4 Introduction ...
Страница 48: ...Safety Tests 36 Performance Verification ...
Страница 72: ...Troubleshooting Guide 60 Chapter 5 Troubleshooting ...
Страница 99: ...Disassembly Procedures A3 Chapter 6 Disassembly Guide 87 NiBP pump assembly Main PCB Main ribbon cable ...
Страница 100: ...Disassembly Procedures A3 88 Chapter 6 Disassembly Guide ...
Страница 102: ...Top Level Assembly 90 Chapter 7 Spare Parts Top Level Assembly Figure 1 A1 Top Level Assembly Drawing 1 of 2 ...
Страница 103: ...Top Level Assembly Chapter 7 Spare Parts 91 Figure 2 A1 Top Level Assembly Drawing 2 of 2 ...
Страница 105: ...Top Level Assembly Chapter 7 Spare Parts 93 Figure 3 A3 Top Level Assembly Drawing ...
Страница 107: ...Front Case Assembly Chapter 7 Spare Parts 95 Front Case Assembly Figure 4 A1 Front Case Assembly Drawing 1 of 3 ...
Страница 108: ...Front Case Assembly 96 Chapter 7 Spare Parts Figure 5 A1 Front Case Assembly Drawing 2 of 3 ...
Страница 110: ...Front Case Assembly 98 Chapter 7 Spare Parts Figure 7 A3 Front Case Assembly Drawing 1 of 2 ...
Страница 111: ...Front Case Assembly Chapter 7 Spare Parts 99 Figure 8 A3 Front Case Assembly Drawing 2 of 2 ...
Страница 114: ...Rear Case Assembly 102 Chapter 7 Spare Parts Figure 9 A3 Rear Case Assembly Drawing 1 of 3 ...
Страница 115: ...Rear Case Assembly Chapter 7 Spare Parts 103 Figure 10 A3 Rear Case Assembly Drawing 2 of 3 ...
Страница 116: ...Rear Case Assembly 104 Chapter 7 Spare Parts Figure 11 A3 Rear Case Assembly Drawing 3 of 3 ...
Страница 124: ...Repacking In A Different Carton 112 Chapter 8 Packing For Shipment ...
Страница 132: ...Theory of Operation and System Architecture 120 Chapter 10 Training Program Block Diagram A1 Figure 1 A1 Block Diagram ...
Страница 144: ...Preventative Maintenance 132 Chapter 10 Training Program ...
Страница 150: ......