Troubleshooting Guide
52
Chapter 5 - Troubleshooting
4. Simultaneously press the
CONTRAST
button
and the
ON/
STANDBY
button
to power up the monitor. Keep the contrast
button depressed until the monitoring screen appears (after 10 seconds).
The error code printout is generated automatically.
If error codes listed on the Diagnostic Mode error code screen or on the error
log printout are in the range from 1 to 65 (hex), a hardware problem has been
detected. Refer to the section titled, “Serviceable Hardware Error Codes” for
additional information on these codes.
Serviceable Hardware Error Codes
Serviceable Hardware
Error Codes for A1
Listed below are error codes for the A1 that correspond to hardware problems,
and the recommended actions to take should such an error be encountered.
Hex Code
Explanation
Recommended Action
1
Improper shutdown.
1. Cycle power.
2. If this error persists, return
monitor for service.
2
NiBP Sensor Error.
• The two pressure system.
transducers do not agree.
1. Check for blocked tubing in
the pneumatic system.
2. Replace Main PCB.
3
NiBP Pressure Violation Error.
• The pressure on the cuff
could not be removed by
normal means.
• A fault has been detected in
the NiBP system that could
not be handled by releasing
pressure by normal means.
1. Cycle power.
2. Check for blocked tubing in
the pneumatic system.
3. Replace Main PCB.
4
The measured value of the 3.3V
power supply is low.
1. Check power supply.
2. Replace Main PCB.
5
The measured value of the 3.3V
power supply is high.
1. Check power supply.
2. Replace Main PCB.
7
The measured value of the 12V
power supply is high.
1. Check power supply.
2. Replace Main PCB.
8
The measured value of the 5V
power supply is low.
1. Check power supply.
2. Replace Main PCB.
9
The measured value of the 5V
power supply is high.
1. Check power supply.
2. Replace Main PCB.
Содержание M3921A
Страница 12: ...xii Contents ...
Страница 16: ...Related Documents 4 Introduction ...
Страница 48: ...Safety Tests 36 Performance Verification ...
Страница 72: ...Troubleshooting Guide 60 Chapter 5 Troubleshooting ...
Страница 99: ...Disassembly Procedures A3 Chapter 6 Disassembly Guide 87 NiBP pump assembly Main PCB Main ribbon cable ...
Страница 100: ...Disassembly Procedures A3 88 Chapter 6 Disassembly Guide ...
Страница 102: ...Top Level Assembly 90 Chapter 7 Spare Parts Top Level Assembly Figure 1 A1 Top Level Assembly Drawing 1 of 2 ...
Страница 103: ...Top Level Assembly Chapter 7 Spare Parts 91 Figure 2 A1 Top Level Assembly Drawing 2 of 2 ...
Страница 105: ...Top Level Assembly Chapter 7 Spare Parts 93 Figure 3 A3 Top Level Assembly Drawing ...
Страница 107: ...Front Case Assembly Chapter 7 Spare Parts 95 Front Case Assembly Figure 4 A1 Front Case Assembly Drawing 1 of 3 ...
Страница 108: ...Front Case Assembly 96 Chapter 7 Spare Parts Figure 5 A1 Front Case Assembly Drawing 2 of 3 ...
Страница 110: ...Front Case Assembly 98 Chapter 7 Spare Parts Figure 7 A3 Front Case Assembly Drawing 1 of 2 ...
Страница 111: ...Front Case Assembly Chapter 7 Spare Parts 99 Figure 8 A3 Front Case Assembly Drawing 2 of 2 ...
Страница 114: ...Rear Case Assembly 102 Chapter 7 Spare Parts Figure 9 A3 Rear Case Assembly Drawing 1 of 3 ...
Страница 115: ...Rear Case Assembly Chapter 7 Spare Parts 103 Figure 10 A3 Rear Case Assembly Drawing 2 of 3 ...
Страница 116: ...Rear Case Assembly 104 Chapter 7 Spare Parts Figure 11 A3 Rear Case Assembly Drawing 3 of 3 ...
Страница 124: ...Repacking In A Different Carton 112 Chapter 8 Packing For Shipment ...
Страница 132: ...Theory of Operation and System Architecture 120 Chapter 10 Training Program Block Diagram A1 Figure 1 A1 Block Diagram ...
Страница 144: ...Preventative Maintenance 132 Chapter 10 Training Program ...
Страница 150: ......