78
UV LAMP PARAMETERS
7.
On the
UV Setting
screen, by holding the
UV 24h
Mode
slide key and sliding it right, the UV 24h
Mode is turned to ON. Press the
Apply
key to start
the UV 24-hour mode. The display returns to the
Tools #1
screen.
8.
The UV lamp lights continuously for 24 hours. “UV : ON” is displayed on the UV lamp condition display
when UV lamp is lighting.
Note:
·
The UV 24-hour mode may cause the automatic set temperature alarm because of a rising chamber
temperature.
·
After procedure
8
, by opening the outer door when UV lamp is lighting, UV lamp is turned OFF and UV
24-hour mode is canceled by opening the outer door. Redo from procedure
4
to start the UV 24-hour mode
again.
9.
Press the
Top
key to return to the
Top
screen.
10.
24 hours after, UV lamp turns OFF automatically. Install all attachments removed in the procedure
1
.