19
In case of double stack
For stacking the incubators surely, refer to the
procedure included with the optional double
stacking bracket MCO-170PS or the stacking plate
MCO-170SB.
Note:
·
Two hooks are attached to the rear of the upper
incubator. When stacking incubators, fix the upper
incubator to the wall with these hooks and wire or
chain (Fig. 4).
·
When 2 or more incubators are performed dry
heat sterilisation simultaneously, ensure that the
capacity of power supply is sufficient.
·
When 2 or more incubators are performed dry
heat sterilisation simultaneously, the surface
temperature may be higher than the case of one
incubator.
·
When stacking the incubators on our CO
2
incubator or O
2
/CO
2
incubator other than this product, use the
stacking plate MCO-170SB. Refer to Table 10 on page 95.
When the incubator is not in use
Empty the water from the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
Before moving the incubator
Before moving the incubator, empty the water from the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the power supply cord will not be damaged. Failure to do so may result
in electric shock or fire.
Hooks
Wire
Fig. 4