38
ALARM PARAMETERS
1.
Press the
Menu
key to lead the
Menu
screen.
2.
Press the
Tools #1
key to lead the
Tools #1
screen.
3.
On the
Tools #1
screen,
·
Press the
Alarm Setting #1
key to lead the
Alarm
Setting #1
screen, it is possible to set automatic set
temperature alarm and automatic set CO
2
density
alarm (go to procedure
4
).
·
Press the
Alarm Setting #2
key to lead the
Alarm
Setting #2
screen, it is possible to set alarm delay,
door alarm delay, ring back and remote alarm (go
to procedure
5
).
4.
On the
Alarm Setting #1
screen, input each
parameter. Press the
Apply
key to save the input
value. The display returns to the
Tools #1
screen.
●
Each parameter setting
·
Temp. Alarm:
When the chamber temperature exceeds the scope,
the set temperature ± the set value of Automatic set
temperature alarm, the alarm is activated. Settable
range: ±1.0
o
C~±5.0
o
C, factory setting: ±1.0
o
C.
·
CO
2
Alarm:
When the chamber CO
2
density exceeds the scope, the set CO
2
density ± the set value of Automatic set
CO
2
density alarm, the alarm is activated. Settable range: ±0.5 %~±5.0 %, factory setting: ±1.0 %.