PERFORMANCE
Product name
CO
2
Incubator MCO-170AICUVDL
Model number
MCO-170AICUVDL-PA
Rated voltage, frequency
AC 110 V-120 V, 60 Hz
During cultivation
Temperature control range
Ambient temp5
o
C to 50
o
C* (ambient temperature: 5
o
C to 35
o
C)
Temperature distribution
±0.25
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
Temperature variation
±0.1
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
CO
2
control range
0 % to 20 %
CO
2
variation
±0.15 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, no load)
Chamber humidity
95 %R.H.±5 %R.H.
Applicable environment
condition
Temperature: 5
o
C to 35
o
C, Humidity: 80 %R.H. max.
(The designed performance may not be obtained
if the ambient temperature is equal or less than 15
o
C,
or if the ambient humidity is high.)
Noise level
25 dB (A scale)
Heater
516 W
Power consumption
Max. 545 W
Heat emission
Max. 1,860 kJ/h
Amperage
Max. 4.8 A
During dry heat sterilisation
Applicable
environment
condition
Temperature: 15
o
C to 30
o
C, Humidity: 80 %R.H. max.
Heater
909 W
Power consumption
Max. 935 W
Heat emission
Max. 3,280 kJ/h
Amperage
Max. 8.0 A
*When set temperature is 37
o
C, ambient temperature must be 32
o
C or less. Regardless of ambient
temperature, the maximum of temperature control range is always 50
o
C.
Note: