TROUBLESHOOTING
Symptom
Items to check and countermeasures
Normal cultures are not possible,
and the CO
2
gas density is
suspect.
Is the ambient air environment around the incubator normal? Is there a source of polluted
gas in the vicinity?
CO
2
gas is not being injected.
The CO
2
control method for the incubator is the ON-OFF method. CO
2
gas is
intermittently injected as the gas density in the chamber approaches the set value.
Injections may be stopped for periods of approximately 15 seconds, but that is not an
error.
The gas is not injected until the temperature of the CO
2
sensor becomes stable enough
approx. 1 hour, after turning ON the power switch or recovering from power failure.
The CO
2
gas density is slow to
recover.
A HEPA filter is used for the incubator CO
2
gas piping. If gas density is slow to recover
when the CO
2
gas pressure is normal, it is possible that the HEPA filter may be clogged.
Contact our sales representative or agent.
Is there little gas remaining in the CO
2
gas cylinder?
Is the secondary pressure for the gas regulator at the specified value of 0.03
MPa(G)
~
0.1 MPa(G) (0.3 kgf/cm
2
(G)
~
1 kgf/cm
2
(G), 4.4 psi(G)
~
14.5 psi(G))?
Is the gas tube blocked?
Is the duct securely attached? Attach the duct properly to the 4 points hooks. (Fig. 2 on
page 24)
Is the fan attached properly? Confirm if the fan is pushed all the way to the motor shaft.
(Fig. 1 on page 24)
UV lamp lights when the outer
door is open.
Does something push the door switch?
The outer door does not open.
When the power switch is OFF, the electric lock is locked and the outer door does not
open. Either turn ON the power switch or use the accessory unlock key to override the
electric lock.
During dry heat sterilisation the outer door is electrically locked and does not open.
Dry heat sterilisation ends in the
middle by an error.
Are silicon caps attached access port?
Is the inner door open?
Note:
If the problem still has not been solved after trying the above checks and countermeasures, or for
any problems not covered here, contact our sales representative or agent.
Keep an electric product which emits an electromagnetic wave away from this product. A noise from an
electromagnetic wave may cause malfunction to this product.