84
STD GAS AUTO CALIBRATION KIT (OPTION)
5.
On the
STD Gas Setting
screen, input CO
2
density of the connected standard CO
2
gas cylinder.
Press the
Apply
key to save the input value. The
display returns to the
Tools #1
screen.
Settable range: 4.0 %~21.0 %.
Factory setting: 5.0 %.
6.
Press the
STD Gas Calibration
key to lead the
STD Gas Calibration
screen.
7.
On the
STD Gas Calibration Step1
screen,
system check starts automatically. If the system is
normal, display leads the
STD Gas Calibration
Step2
screen.
8.
On the
STD Gas Calibration Step2
screen, press
the
OK
key to lead the
STD Gas Calibration Step3
screen.