39
5.
On the
Alarm Setting #2
screen, it is possible to
set each alarm. Press the
Apply
key to save the
input value and setup. The display returns to the
Tools #1
screen.
●
Each setting
·
Alarm Delay:
The function is that when the incubator is in the alarm state of Automatic set temperature or of Automatic
set CO
2
density, the alarm buzzer will sound after the set time of alarm delay passed.
Settable range: 0 minute~15 minutes, factory setting: 15 minutes.
Note:
When the incubator is recovered from the alarm state within the set time of alarm delay, the buzzer
doesn’t sound after the elapse of the set time of alarm delay.
·
Door Delay:
The function is that when the incubator is in the alarm state of door, the alarm buzzer will sound after the
set time of door alarm delay passed.
Settable range: 1 minute~30 minutes, factory setting: 2 minutes.
Note:
When the incubator is recovered from the alarm state within the set time of door alarm delay, the
buzzer doesn’t sound after the elapse of the set time of door alarm delay.
·
Ring Back:
The function is that the alarm buzzer sounds again when the alarm state still continues after the set time of
ring back passed even though the alarm buzzer was stopped by pressing the
Buzzer
key. By holding and
sliding the
Ring Back
slide key to the right, the Ring Back is turned to ON.
Settable range: 1 minute~99 minutes, factory setting: 30 minutes.
Note:
In the case of Err01 (CO
2
gas cylinder empty), Err11•12 (CO
2
sensor error) or Door alarm, the alarm
is not re-activated because the alarm itself is canceled by pressing the
Buzzer
key. Incidentally in the case
of Err18 (UV lamp failure), the alarm is not re-activated if the
Buzzer
key is pressed after the UV lamp ON
period elapses (refer to page 87~88).
·
Remote Alarm:
The function is that the remote alarm is continued even though the buzzer is stopped by pressing the
Buzzer
key. By holding and sliding the
Remote Alarm
slide key to the right, the Remote Alarm is turned to
ON (not in conjunction with the
Buzzer
key). Factory setting: ON.
6.
(From procedure
4
and
5
) Press the
Top
key to return to the
Top
screen.