77
Lighting UV lamp for 24 hours
If the chamber has been contaminated by dirt or by spilling the medium, use the following procedure to
disinfect the chamber by lighting the UV lamp for 24 hours.
1.
Remove attachments from the chamber, including the trays, the fan cover, the duct, the fan, the
humidifying pan, and the humidifying pan cover. If possilbe, autoclave them for sterilisation, otherwise
clean and wipe off them with alcohol for disinfection.
2.
Clean and wipe off inside the chamber with alcohol for disinfection.
3.
Set the CO
2
density to 0 %. Refer to page 32~33.
4.
Press the
Menu
key to lead the
Menu
screen.
5.
Press the
Tools #1
key to lead the
Tools #1
screen.
6.
Press the
UV Setting
key to lead the
UV Setting
screen.