42
OPERATION/ALARM/STERILISATION LOG
4.
On the
Setting
screen, input Log Interval. Press
the
Apply
key to save the input value. The display
returns to the
Log
screen.
Settable range: 2 minutes~30 minutes.
Factory setting: 6 minutes.
Note:
It is possible to register 8-digit alphanumeric
characters as the Unique ID. Refer to page 48.
Note:
Relation between log interval and the estimated amount of data that can be saved
Log interval=2 minutes: Approx. 46 days
Log interval=6 minutes: Approx. 135 days
Log interval=30 minutes: Approx. 664 days
When saving data more than the above, and the data is overwritten and the old data is delated.
5.
Press the
Top
key to return to the
Top
screen.
Displaying operation log
Operation log saved in the incubator can be displayed graphically on the LCD touch panel.
1.
Press the
Menu
key to lead the
Menu
screen.
2.
Press the
Log
key to lead the
Log
screen.